MTF Tool
Designed for the measurement of the spatial resolution of digital breast tomography and mammography systems.
Features
The MTF Tool is designed for the measurement of the spatial resolution of radiology and mammography systems. The MTF Tool is delivered with an aluminum plate.
TEST ITEMS
MTF tool of Stainless steel: The overall dimensions of the MTF tool are: 120 x 60 x 0.8 mm. Both edges of 120 lengths have a ripple of ± 50 micrometers perpendicular to the surface of the plate.
Aluminum Plate: The overall dimensions of the Aluminum Plate are: 200 x 200 x 2 mm. Thickness tolerance ± 50 micrometer
All specifications
COMPATIBILITY
The 4th edition of the European Guidelines for Quality Assurance in Breast Cancer Screening and Diagnosis (EPQC)
Euref Type test protocol version 1.3
IEC 61223-3-2:2007
NHSBSP Equipment Report 0604, Commissioning and Routine Testing of Full Field Digital Mammography Systems Published April 2009 (Version 3)
IEC 61223-2-6:2019
EUREF Breast Tomosynthesis Quality Control Protocol
Specifications are subject to change without notice. Always read product labelling/inserts for complete instructions, warnings and restrictions.